Contact

Products

Fluoride ion measuring unit (PWCA-F)

✔ Compact design fluoride ion measuring device that can be easily added to the existing one

Features

● Measurement is possible by simply drawing in sample water from existing filtration equipment, etc.
● Equipped with a sensor holder and stirrer that are convenient for calibration (easy to calibrate with standard solution)
● Fluoride ion measurement can be added up to 2 locations by adding a fluoride ion measurement unit.
● Continuous measurement and measurement value storage function by SD card (built-in data logger function)

Use

● Measurement of “fluoride ion concentration” contained in factory effluent
● Plating / stainless steel pickling factory (using hydrofluoric acid / mixed acid)
● For monitoring the final effluent (within the standard)

Fluorine is used in various applications such as cleaning processes for semiconductor manufacturing and glass melting accelerators, but it does not biodegrade due to its high stability, and there is concern about toxicity to the human body, so strict emission standards are national. It is more defined.

Fluorine and its compounds What is discharged to public water areas other than sea areas: 8mg F/L
What is discharged into the sea area: 15mg F/L

specification

Product name Panel type fluoride ion measurement monitoring unit PWCA-F
Model PWCa F00010C0000EN01
Measurement range Fluoride ion F-
0.05 to 10.0 mg / L / 0.5 to 100 mg / L (water temperature 0 to 35 ° C)
Measured value output (0) 4-20mA DC x 1 point (with scaling function)
Calibration function 1-point calibration or 2-point calibration
Calibrate with standard solution, photometer, etc.
Alarm output Upper and lower limits a / b contacts x 2 points (with hysteresis function)
display English notation
Power supply 100-240VAC 50/60Hz
Ambient temperature 0~40℃
Installation method Indoor installation / Steel plate box (optional) installed for outdoor
Sample water 0.2-3L / min (350mL / min recommended, water pressure 0.1MPa or less)

<Fluoride ion (F-) measurement method (fluoride lantern electrode method)>
This is a measurement method used in JIS, etc., which uses a fluorine ion electrode that uses a lanthanum fluoride solid film that responds to fluorine ions.

<Interfering substance>
OH-Al3 + Si4 + Ca2 + HF (hydrofluoric acid) does not interfere (cannot be measured)

Head Office

1-12-11 Tagawakita, Yodogawa-ku,Osaka
532-0021

Overseas Business Department

TEL +81-6-6301-6460
FAX +81-6-6308-3022